DAkkS (17025) calibration resistance thermometer comparison calibration
-196 °C to 961 °C

DAkkS (17025) comparison calibration of a resistance thermometer in the range from -196 °C up to 961 °C.

Description

DAkkS (17025) comparison calibration

DAkkS (17025) calibration of a resistance thermometer in a calibration bath or calibration furnace, in a nitrogen cryostat or on boiling nitrogen at 5 temperature points (distributed over the calibration range or according to customer requirements). The following measurement uncertainties can be achieved (depending on the test unit):

Calibration range Measurement uncertainty Calibration method
-196 °C 15 mK Nitrogen boiling point
-180 °C to -80 °C 60 mK* Nitrogen-Cryostat
-80 °C to -60 °C 15 mK
Calibration bath
  
-60 °C to 90 °C 10 mK
90 °C to 200 °C 30 mK
200 °C to 500 °C 40 mK
Calibration furnace

500 °C to 660 °C 50 mK
660 °C to 961 °C 150 mK

*DAkkS (17025) accreditation procedure in progress

Before calibration, the resistance thermometer is annealed. The self-heating and the hysteresis of the resistance thermometer are tested before the start of calibration and taken into account when calculating the calibration results.

Calibration time: approx. 5 working days or by arrangement.

 