Pt/Pd
precision thermocouple
(0 °C to 1500 °C)

4.307,00 6.745,00 

Each eXacal thermometer is supplied in a matching case. Optionally, a high-quality wooden carrying case can be ordered.

Manufacturer: Klasmeier / eXacal
Model: Pt/Pd Thermocouple

Description

Platinum / Palladium thermocouples (Pt/Pd) are the most precise thermocouples manufactured in the eXacal manufactory.

  • Temperature range: 0 °C to 1500 °C
  • stress compensating design
    The design of the measuring point allows stress-free expansion of the thermocouple wires.
  • Excellent inhomogeneity
    The high purity of 99.99% and the thermal annealing of the thermowires ensure excellent inhomogeneity.
  • Small measurement uncertainty and low drift
    The achievable measurement uncertainties are better by a factor of 2 than with type R/S thermocouples.

Pt/Pd thermocouple can be used in the temperature range from 0 °C to 1500 °C with very small measurement uncertainties. As a special feature, this thermocouple uses a stress-compensating junction of the thermocouple wires. This special design ensures stress-free expansion of the platinum and palladium thermocouple wires. The different thermal expansion coefficients of the two thermocouple wires are thus compensated without mechanical stresses changing the Seebeck coefficient.

Uncertainty of measurement

 

Questions about the thermometer?



    Additional information

    Select calibration

    inclusive DAkkS Calibration, without DAkkS Calibration

    Technical Data

    Model No. KK-150
    Connection Stress free copper wire
    Type Platinum / Palladium (Pt/Pd)
    Temperature range 0 °C to 1500 °C
    Thermowire
    • Platinum / Palladium (Pt/Pd)
    • 99,99 % Purity
    • thermally pretreated and aged
    Measuring point
    • made of gas-tight high-purity ceramic
    • Stress-free expansion of the thermocouple wires possible
    Characeristic IEC 62460
    Measurement uncertainties (at a DAkkS calibration)
    Fixed point calibration
    0,01 °C
    419,527 °C
    660,323 °C
    961,78 °C
     0,2 K
     0,2 K
     0,2 K
     0,2 K
    Comparison calibration
    0 °C to 962 °C
    >962 °C to 1200 °C
     0,4 K
     1,5 K
    Dimensions
    Protection tube measuring point Ø 7 mm x 500 mm
    Handle Ø 20 mm x 95 mm
    Length Thermowire 2 m
    Cold junction Ø 6 mm x 280 mm (Stainless stell)
    with two meters connection cable
    made of low EMF copper
    Length Connection cable 2 m

    Cart